Manual Calibration
-
Ensure the Mini-OX is configured for manual calibration. Refer to Setting the Output Type section.
-
Place the sensor probe in the calibration gas, typically fresh air.
-
Allow the output tostabilize for at least 5 minutes (10 minutes if powering from cold).
-
Referring to page 9; Apply GND to the CALIBRATE input (pin 3) for a minimum 5s or until the CYCLE output and green LED blink at a steady 1Hz. Remove GND from pin 3. Manual Calibration is now initialised.
-
Adjust the MANUAL CAL POT until the output equals the correct value of the calibration gas concentration (see page 18).
-
Re-apply GND to pin 3 for a minimum 5s. During the 5s the CYCLE output/LED will blink rapidly, go high/on, go low/off then return to cycling normally to indicate normal operation has resumed. At this point remove GND from pin 3.
NOTE:The output will now track to the correct value for the calibration gasc.
-
Calibration is complete.
NOTE: New calibration values are stored in memory and retained on power loss.